AichromBond-1 C18 |
¡ñ strongly hydrophobic ( among the most hydrophobic phases)
¡ñ Highest inertness, method development optimization
¡ñ Good peak shape for basic, acidic and neutral compounds
¡ñ Better high volume injection tolerance ( peaks on Luna start to broadening at 20 ul inj)
¡ñ High performance over a wide pH range¡ª1.5~10.0
¡ñ Particle size from 2.1um to 4.6um
¡ñ Long lifetime with extra dense bonding and double endcapping
¡ñsimilar selectivity to Luna C18, and even closer to Supeco Discovery C-18
|
|
Column Specifications |
| Bonded Phase |
Carbon Load |
Pore Size |
Surface Area |
pH Range |
Endcapped |
| AichromBond-1 |
22% |
100A |
380 m2/g |
1.5~10.0 |
Double |
|
|
| |
Order Information |
| Description |
Size(mm) |
Particle Size(um) |
P/N |
Analytical
|
150mm¡Á4.6 |
5 |
1805071-150 |
| Analytical |
250mm¡Á4.6 |
5 |
1805071-250 |
| Narrow Bore |
150mm¡Á2.1 |
5 |
1805072-150 |
| Narrow Bore |
250mm¡Á2.1 |
5 |
1805072-250 |
|
|
| |
Performance Comparison: |
¡ñ Large volume injection and least peak tailing for bases

|
 |
| |
 |
| |
Sample: Uracil, 2,3-dihydroxynaphtlene, Propanolol, Naphtlene, Amitriptyline
Column: AichromBond-1 C18, 4.6 x 150 mm, 5 um particle size
Mobile phase: 10 mm sodium phosphate buffer (pH=7)/AC N ( 25:75)
Temperature: 30 oC
Detection: UV 254 nm
Flow: 1 mL/min. |
|
| |
| ¡ñ Low Pressure |
 |
| |
 |
Sample: 1. Uracil, 2. 2,3 dihydroxy naphtlene, 3. Propanolol, 4. Naphtlene, 5. Amitritypline
Injection: 5 uL
Column: 4.6 mm x 150 mm, 5 um
Mobile phase: 10 mmol sodium phosphate buffer (pH=7)/Methanol = 15:85
Flow: 1 mL/min
Temperature: 30 oC
Detection: UV 254 nm |
| |
AichromBond-AQ C18 |
¡ñ Inertness: it is one of most base-friendly columns, comparable with Xterra and even better than Inertsil, is far better than Kromasil and Hypersil.
¡ñ Efficiency: the best for polar compounds at any pH.
¡ñ Large volume injection: maintain very high efficiency even if the injection volume is exceptional large.
¡ñ 100% water compatible: very unique for a universal C18 column; far better peak shape, retention and efficiency than those so-called high aqueous ¡°AQ¡± columns.
¡ñ Low pH Stability (pH=1.5): next to the Zorbax SB C18, it has better stability than most of popular brand columns on the market, and is far better than any domestic column at low pH. |
|
Column Specifications |
| Bonded Phase |
Carbon Load |
Pore Size |
Surface Area |
pH Range |
Endcapped |
| AichromBond-AQ |
19% |
100A |
380 m2/g |
1.5~10.0 |
Single |
|
|
| |
Order Information |
| Description |
Size(mm) |
Particle Size(um) |
P/N |
Analytical
|
150mm¡Á4.6 |
5 |
1805AQ1-150 |
| Analytical |
250mm¡Á4.6 |
5 |
1805AQ1-250 |
| Narrow Bore |
150mm¡Á2.1 |
5 |
1805AQ2-150 |
| Narrow Bore |
250mm¡Á2.1 |
5 |
1805AQ2-250 |
|
|
| |
Performance Comparison: |
| ¡ñ Inertness and efficiency comparison |
 |
Sample: uracil, toluene, doxepin , amitriptyline
Column: AichromBond AQ C18, 4.6 x 150 mm, 5 um particle size
Mobile phase: 10 mM sodium phosphate (pH=7) in 60% ACN
Flow: 1 mL/min.
Temperature: 30¡æ |
|
| ¡ñ Large volume injection |
|
Large Volume Injection (injection volume =100 uL, )
Sample: 4-methoxybenzenesulfonamide, 10 ug/mL in MeOH/Water (1:1)
Column: AichromBond AQ C18, 4.6 x 150 mm, 5 um particle size
Mobile phase: MeOH/Water
Flow: 1mL/min.
Temperature: 30¡æ |
| |
| ¡ñ Compatible to 100 water mobile phase |
 |
Sample: Uridine
Column: AichromBond AQ C18, 4.6 x 150 mm, 5 um
Column dimension: 4.6 x 150 mm
Mobile phase: 100% water
Flow: 1mL/min.; The flow was stopped once for 5 minutes during the testing period for each column.
Temperature: 30¡æ |
| |
| ¡ñ Stability at pH=1.5 |
 |
Sample: Uracil, Doxepin, Amitriptyline, Toluene.
Column: AichromBond AQ C18, 4.6 x 150 mm, 5 um
Mobile phase: TFA (0.25%, pH=1.5) in 60% acetonitrile
Flow: 1 mL/ min
Temp: 40¡æ |
| |